5028i Continuous Particulate Monitor

Conduct simultaneous, real-time measurement of PM10 and PM2.5 and benefit from lower maintenance requirements with the Thermo Scientific™ Model 5028i Continuous Particulate Monitor. The unit combines dual channel sampling technology and beta attenuation analysis on the proven iSeries platform to measure PM10 and PM2.5 simultaneously. The Model 5028i offers continuous, non-stepwise measurement of particulates and complies with U.S. EPA PM-10 (EQPM1102-150) and PM-2.5 (EQPM0609-183).

Key Features:

The Model 5028i Continuous Particulate Monitor combines dual channel sampling technology and beta attenuation analysis on the proven iSeries platform to measure PM10 and PM2.5 simultaneously. The Model 5028i monitor offers continuous, non-stepwise measurement of particulates. The key features of this state-of-the-art monitor are:

  • U.S. EPA Approved PM-10 (EQPM1102-150) and PM-2.5 (EQPM0609-183)
  • Federal Equivalent Method (FEM)
  • Continuous, non-stepwise measurement of PM-10 and PM-2.5 simultaneously
  • Internal backup power to reduce reliable data output time after power off/interruption
  • Filter tape-based monitor, with increased tape volume for longer unattended operation
  • New chassis design to make maintenance and service much easier

Description

Specification

FuenteCarbon-14, < 3.7 MBq (<100 µCi)
Measuring range0 a 0,1 mg/m3, 1,0, 2,0, 3,0, 5,0, 10,0 mg/m3; 0 a 100,1,000, 2,000, 3,000, 5,000, 10,000 ug/m3
Minimum Detection Limit< 4 μg/m3 (1 hour) @ 2σ; < 1 μg/m3 (24-hour) @ 2σ
Resolution0,1 μg/m3
Precision± 3.0 μg/m³ < 80 μg/m³; 4-5 μg/m³ > 80 μg/m³ (24-hour)
RMS AccuracyPM2.5 < 5%, PM10 < 5% (24-hour)
Accuracy (for mass measurement)±5% using a set of NIST traceable mass sheets
Air Flow Range1 m3 /h (16.67L/min) measured through an internal subsonic orifice
Precision Sample Flow±2% of measured value
Accuracy Sample Flow< 5% of measured value
Mass Concentration60 to 3,600 seconds and 24-hour
Data Output RateEvery 1 second
Operating temperatureThe temperature of the sampled air can vary between -30 and 45°C. 5028i units must be protected against the climate within the range of 4°C to 50°C. An optional full outer cover provides complete weather protection”
No condensation< 95% RH entro del monitor 5028i
SalidaVoltage selectable, RS232/RS485, TCP/IP, 10 relays status and power failure indication (standard). output of isolated current 0-20 or 4-20 mA (optional)
Prohibited16 digital inputs (standard), eight analog inputs 0 to 10 VDC (optional), 8 user-defined analog outputs user (0 -1 or 0 -5 VDC)
Energy Requirements110VAC-120VAC, 220-240VAC, 50/60Hz 700 watts (110V); 700 watts (220V) Maximum
Bomba220 VAC 50/60 Hz, 2.0 A
Physical Dimensions18.96” (48.2 cm) W × 22.23” (56.5 cm) D × 23.43” (59.5 cm) H
Weight110VAC: 77 lbs (35 kg), 220VAC: 66 lbs (30kg)
ProtocolsC-Link, MODBUS, Gesytec (Bayern-Hessen), protocol ESM, data transmission and NTP protocols network time). Simultaneous connections from different locations over Ethernet
Electrical Designations and Security“Designed to comply with CE: EN 61326: 1997 + A1: 1998 +A2: 2001 + A3: 2003, EN: 61010-1; UL:61010-1:2004; CSA: C22.2 No. 61010-1: 2004; FCC: Part 15 Subpart B, Class B”
Approvals and CertificationsU.S. EPA PM-10 Monitor Equivalent: EQPM1102-150; US EPA PM-2.5 Monitor Equivalent: EQPM0609-183