48 i CO Analyzer
Measure the amount of carbon monoxide (CO) in ambient air with the Thermo Scientific™ Model 48i CO Analyzer. The gas filter correlation technique as utilized by the Model 48i gas filter correlation analyzer offers over 30 years of field-proven performance in providing accurate and reliable readings.
The Model 48i is based on the principle that carbon monoxide (CO) absorbs infrared radiation at a wavelength of 4.6 microns. Because infrared absorption is a nonlinear measurement technique, it is necessary for the instrument electronics to transform the basic analyzer signal into a linear output. The 48i uses an exact calibration curve to accurately linearize the instrument output over any range up to a concentration of 10,000ppm.
Key Features:
- Gas filter correlation technology
- Linearity through all ranges
- Measures concentrations up to 10,000ppm
- Standard dual range and auto ranges
- Temperature and pressure correction
iSeries Advantage:
- Ethernet connectivity for efficient remote access
- Enhanced user interface with one button programming and large display screen
- Flash memory for increased data storage and user downloadable software
- Enhanced electronics design optimizes product commonality
- Improved layout for easier accessibility to components
The Mode 48i CO Analyzer is also available with an oxygen sensor module.
The Model 48i with Oxygen uses paramagnetic sensor technology to extend CO measurement capabilities, reading and reporting the concentration of oxygen in the sample stream up to 100%. Additionally, programming allows the user to correct the CO reading for the amount of oxygen in the sample.
Preset ranges | 0-1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000 and 10000 ppm 0-1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000 and 10000 mg/m3 |
Custom ranges | 0-1 to 10000 ppm, 0-1 to 10000 mg/m3 |
Zero noise | 0.02 ppm RMS (30 second averaging time) |
Lower detectable limit | 0.04 ppm |
Zero drift | < 0.1 ppm |
Span drift | +/-1% full scale |
Response time | 60 seconds (30 second average time) |
Precision | +/-0.1 ppm |
Linearity | +/-1% full scale < 1000 ppm +/-2.5% full scale > 1000 ppm |
Sample flow rate | 0.5-2 liters/min. |
Interferences | < lower detectable limit except for the following: (EPA Levels) NO < 3 ppb, M-Xylene < 1 ppb, H2O < 3% of reading |
Operating temperature | Performance specifications based on operation within 20°-30°C range (per U.S. EPA Guidelines). Instrument may be safely operated over the range of 0°- 45°C. |
Power requirements | 100 VAC, 115 VAC, 220-240 VAC +/-10% @ 275W |
Size and weight | 116.75″ (W) × 8.62″ (H) × 23″ (D), 49 lbs. (22.2 kg) |
Outputs | Selectable voltage, RS232/RS485, TCP/IP, 10 status relays and power fail indication (standard) 0-20 or 4-20 mA isolated current output (optional) |
Inputs | 16 digital inputs (standard), 8 0-10Vdc analog inputs (optional) |
Approvals and certifications | U.S. EPA Equivalent Method: RFCA-0981-054 MCERTS Certified: Sira MC070095/00 EN14626: TÜV 936/21203248/A Report UKCA |