oxi::lyser

The oxi::lyser is an optical multi-parameter probe that measures the concentration of dissolved oxygen and the temperature directly in the water. It does not need a minimum flow to produce accurate readings and integrates the temperature measurement for On-Line correction. Minimal conventional calibration and virtually zero maintenance.

oxi::lyser™ II monitors dissolved oxygen & temperature

  • s::can plug & measure 
  • Measuring principle: optical / fluorescence
  • Multiparameter sensor
  • Ideal for surface water, groundwater, and drinking water, also wastewater
  • Long term stable and maintenance-free in operation
  • Factory precalibrated
  • Automatic cleaning with compressed air
  • Mounting and measurement directly in the media (InSitu) or in Bypass
  • No flow necessary
  • Operation via s::can terminals & s::can software
  • Minimal maintenance (no consumables)
  • Multiparameter: Measures Dissolved Oxygen (DO) and Temperature
  • Technology: Optical/fluorescence, non-consumable method
  • Maintenance: Virtually zero; no consumables required
  • Installation: Plug & Measure for easy setup
  • Calibration: Factory pre-calibrated, ready to use
  • Performance: Accurate, stable, and long-term reliable
  • Applications: Surface water, groundwater, drinking water, wastewater
 
 
  • Environmental Monitoring: Detect anaerobic conditions in surface and groundwater to protect aquatic life.
  • Drinking Water Treatment: Early warning for microbial or chemical contamination in raw water.
  • Wastewater Treatment: Monitor treatment efficiency and ensure regulatory compliance.
  • Industrial Processes: Control water quality in pre-treatment and effluent monitoring for process consistency.
  • Aquaculture: Prevent anaerobic conditions in fish farms to protect aquatic organisms.
  • Measuring Principle: Optical/fluorescence quenching
  • Parameters: Dissolved Oxygen (O₂) and Temperature
  • O₂ Range: 0–25 mg/L (0–200% saturation)
  • O₂ Resolution: 0.01 mg/L
  • O₂ Accuracy: ±0.02 mg/L or ±1%
  • Temperature Range: 0–60 °C (operating), 0–50 °C (sensor)
  • Temperature Resolution: 0.2 °C
  • Power: 6–16 VDC; Max 0.32 W
  • Interface: RS485 via s::can terminals (IP68)
  • Dimensions: Ø 51 × 160 mm; Weight: 540 g
  • Operating Pressure: 0–6.8 bar
  • Cleaning: Optional automatic compressed air (2–4.5 bar)
  • Housing: CPVC, stainless steel, epoxy; Protection: IP68
Scroll to Top