chlodi::lyser

The chlodi::lyser is an amperometric sensor for monitoring chlorine dioxide. It is ideal for all kinds of water treatment and produces stable readings even at high fluctuations of pH, temperature, and flow. Strong surfactants are tolerated and the chlodi::lyser is not cross-sensitive to chlorine.

 

chlodi::lyser measures chlorine dioxide

  • s::can plug & measure
  • Measuring principle: amperometric (membrane covered)
  • Ideal for all kinds of water treatment
  • Long-term stable and lowest maintenance in operation
  • Replacement of membrane only once a year
  • Readings are stable even at high fluctuations of pH, temperature, and flow
  • Strong surfactants are tolerated
  • Not cross sensitive to chlorine
  • Factory precalibrated
  • Mounting and measurement in a flow cell
  • Operation via s::can terminals & s::can software
  • Additionally also measures temperature
  • Drinking Water Treatment: Ensure safe, high-quality disinfected water.
  • Wastewater Treatment: Monitor effluent quality before discharge.
  • Industrial Processes: Support quality control and process monitoring.
  • Environmental Monitoring: Long-term tracking of surface water and groundwater.
  • Measuring Principle: Amperometric (2-electrode, membrane-covered)
  • Parameters: Chlorine Dioxide (ClO₂) and Temperature
  • Range: 0–2 mg/L (standard model E-528-1-000)
  • Resolution: 0.001 mg/L
  • Response Time: ~1 minute (T₉₀)
  • Operating Temp: 0–50 °C (water), 0–45 °C (ambient); Temp sensor: 5–55 °C
  • Operating Pressure: 0–1 bar
  • Power: 9–30 VDC; typical 0.5 W, max 0.6 W
  • Interface: RS485 via s::can sys plug (IP67)
  • Housing: PVC / Stainless Steel 1.4571
  • Dimensions: Ø 35 × 208 mm; Weight: ≥150 g
  • Recommended Flow: 15–30 L/h (in s::can flow cell)
Scroll to Top