- Wastewater Treatment Plants: Ideal for continuous monitoring of COD, BOD, TOC, and TSS, where heavy organic fouling is common and automatic cleaning ensures reliable performance.
- Industrial Effluent Monitoring: Enables long-term tracking of pollutants and organic loads in industrial discharge streams, reducing maintenance in harsh environments.
- Drinking Water Treatment: Supports stable, reliable monitoring of parameters such as nitrate and UV254 at critical points, even where moderate fouling may occur.
- Environmental Monitoring: Suitable for long-term deployment in rivers, lakes, and groundwater stations that require dependable, low-maintenance operation.

Path length 100 mm Flow assembly autobrush – for spectro::lyser™ path length 100 mm
- For easy and correct bypass installation of s::can spectrometer probes
- For applications with periodic, automatic cleaning
- Cleaning of the measurement window using a rotating brush without dismantling the spectrometer probe
- Purpose: Provides simple and correct bypass (flow-through) installation for the spectro::lyser™ probe.
- Automatic Cleaning: Equipped with an integrated rotating brush for periodic cleaning of the optical windows, preventing fouling without removing the probe.
- Path Length Compatibility: Designed for spectro::lyser™ probes with a 100 mm optical path length, used for higher-sensitivity measurements.
- Material: Constructed from durable POM-C for long-term performance in water quality applications.
- System Compatibility: Works with Badger Meter’s spectro::lyser probes across multiple applications, including:
- Wastewater treatment
- Drinking water treatment
- Environmental and surface water monitoring
- Industrial water and process monitoring
Technical data
- Power supply: 12 VDC
- Assembly: ex works
- Housing material: POM-C
- Dimensions (WxHxD): 74 x 196 x 153 mm
- Weight (min.): 1.7 kg
- Process connection: G 1/4″
- Installation/assembly: Flow-through
- Fitting Temperature range: 0 … 40 °C
- Application pressure: 0 … 6 bar
- Protection class: IP 66
- Article number: F-446-2