Description
Specifications
Preset ranges | 0-0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, and 20 ppm 0-0.1, 0.2, 0.5, 1, 2, 5, 10, 20, and 30 mg/m3 |
Extended ranges | 0-0.2, 0.5, 1, 2, 5, 10, 20, 50, and 100 ppm 0-0.5, 1, 2, 5, 10, 20, 50, 100, and 150 mg/m3 |
Custom ranges | 0-0.05 to 100 ppm 0-0.1 to 150 mg/m3 |
Zero noise | 0.50 ppb RMS (120 second averaging time) |
Lower detectable limit | 1.0 ppb (120 second averaging time) |
Zero drift (24 hour) | < 1 ppb |
Span drift (24 hour) | +/-1% full scale |
Response time | 120 seconds (10 second averaging time) |
Precision | +/-0.4 ppb (500 ppb range) |
Linearity | +/-1% full scale |
Sample flow rate | 0.6 liters/min. |
Operating temperature | 59° – 95° F (15° – 35°C) |
Power requirements | 100 vac, 115 vac, 220-240 vac +/-10% @ 50/60hz, 300W (analyzer) 600W (converter) |
Size and weight | 16.75″ (W) ´ 8.62″ (H) ´ 23″ (D), 60 lbs. (28 kg) Analyzer: 29 lbs. (14 kg) converter |
Outputs | Selectable voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail Indication (standard). 0-20 or 4-20 mA Isolated Current Output (optional) |
Inputs | 16 Digital inputs (standard), 8 0-10Vdc analog inputs (optional) |
Available Options | Tefl on particulate filter, ozone particulate filter, rack mounts, rear extender |